Buckled Micro-Transesophageal Echocardiography Probe During Patent Foramen Ovale Closure

Tutkimustuotos: ArtikkelijulkaisuArtikkeliTieteellinenvertaisarvioitu

Abstrakti

As transcatheter cardiac treatment options continue to increase in number, most hospitals are forced to find solutions for sparse anaesthesia resources. Conscious sedation with the help of a micro-transoesophageal echocardiography (micro TEE) probe can be considered in some procedures. However, new methods can induce unexpected problems. Here, we report 2 cases of buckled micro TEE probes under patent foramen oval closure. Fluoroscopy revealed different mechanisms of buckling in our patients. Esophageal buckling was solved by advancing the micro TEE probe to the ventricle, whereas pharyngeal buckling was solved by careful retracting and straightening of the micro TEE probe in the pharynx.

Alkuperäiskielienglanti
Artikkelie230031
LehtiAnnals of Internal Medicine Clinical Cases
Vuosikerta2
Numero5
Sivumäärä3
DOI - pysyväislinkit
TilaJulkaistu - toukok. 2023
OKM-julkaisutyyppiA1 Alkuperäisartikkeli tieteellisessä aikakauslehdessä, vertaisarvioitu

Lisätietoja

Publisher Copyright:
© 2023 Authors. Published in partnership by the American College of Physicians and American Heart Association.